title: DNA Sequencer Consumables and Service for 3500 | ELIS.AL
url: https://elis.al/
language: en
updated: 2026-10-26
source: ELIS.AL (Euro Lab Instrument Services)
---

# DNA sequencer consumables and service for the 3500 series

New and regenerated capillary arrays, POP polymers and buffer containers for 3500 and 3500xL genetic analyzers, plus consumables, maintenance and repair for the 3130, 3730, SeqStudio and thermal cyclers.

 [Request a quote](https://elis.al/contact/#quote) [Online shop](https://eshop.elis.al)

- **Regenerated arrays**A world first for the 3500, with a reprogrammed RFID tag
- **8 and 24 capillaries**Arrays for the 3500 and 3500xL in 36 cm and 50 cm
- **POP-4, POP-6, POP-7**Polymer pouches with 96, 384 or 960 samples
- **Four regions**Supplying labs in Europe, Asia, the Middle East and Latin America

## Regenerated capillary arrays, a world first for the 3500

We clean and regenerate used capillary arrays for the 3500, 3500xL, 3130 and 3130xl, and reprogram the RFID tag of 3500-series arrays, for much less than the price of a new array. ELIS was the first company in the world to do this for the 3500 and, as far as we know, is still the only one.

- **Cleaned, regenerated, RFID reprogrammed**Your instrument accepts the array and counts its injections again.
- **Much lower price**An original array for much less than a new one, and one array fewer thrown away.
- **Online regeneration**We send the washing liquid and a new RFID tag. Our engineers guide you online, or you follow our protocol.

 [See regenerated arrays](https://elis.al/regenerated-capillary-arrays/) [Request a quote](https://elis.al/contact/?product=reg-3500-36#quote)

![Close-up of a 16-capillary array installed in an ABI PRISM 3100 genetic analyzer, with the capillaries running to the detection cell](https://elis.al/assets/photos/array-800.jpg)

A 16-capillary, 36 cm array (part 4315931) in an ABI PRISM 3100. The 3130xl uses the same array. Photo: [Mark Pellegrini](https://commons.wikimedia.org/wiki/File:ABI_PRISM_3100_Genetic_Analyzer_1.jpg), [CC BY-SA 3.0](https://creativecommons.org/licenses/by-sa/3.0/), cropped

## Consumables for 3500 and 3500xL genetic analyzers

Each item is listed next to the OEM part number your lab orders today, so matching your stock takes seconds.

-

### [Capillary arrays](https://elis.al/capillary-arrays/)

8-capillary arrays for the 3500 and 24-capillary arrays for the 3500xL, for sequencing and fragment analysis.

- **Lengths:** 36 cm, 50 cm
- **OEM refs:** `4404683`, `4404685`, `4404687`, `4404689`
-

### [POP-7, POP-6 and POP-4 polymer](https://elis.al/pop-polymer/)

POP-7 for sequencing and fragment analysis, POP-6 for high-resolution sequencing, POP-4 for human identification.

- **Pouches:** 96, 384, 960 samples
- **On instrument:** Up to 14 days at ≤ 25 °C
-

### [Buffer containers and reagents](https://elis.al/buffers-reagents/)

Anode and cathode buffer containers, and the conditioning reagent that primes and washes the polymer pump.

- **On instrument:** Up to 14 days
- **Storage:** 2–8 °C until use

## Instruments we support

Consumables, maintenance and repair for the capillary electrophoresis systems and thermal cyclers found in most genetics laboratories.

-

### [3500 and 3500xL](https://elis.al/3500-genetic-analyzer/)

New and regenerated arrays, POP pouches and buffers, including the Dx versions.

-

### [3130 and 3130xl](https://elis.al/3130-genetic-analyzer/)

Arrays, regenerated arrays, polymer bottles and running buffer.

-

### [3730 and 3730xl](https://elis.al/3730-dna-analyzer/)

48- and 96-capillary arrays, POP-7, running buffer and online array regeneration.

-

### [SeqStudio Flex](https://elis.al/seqstudio-flex/)

Arrays for the 8 Flex and 24 Flex, with pouches and buffers shared with the 3500.

-

### [SeqStudio](https://elis.al/seqstudio/)

Cartridge v2, cathode buffer containers and service for the 4-capillary benchtop.

-

### [Thermal cyclers](https://elis.al/thermal-cyclers/)

Temperature verification, maintenance and repair for Veriti, ProFlex, 9700 and more.

## Maintenance and repair for genetic analyzers and PCR instruments

Preventive maintenance, spatial and spectral calibration and troubleshooting for capillary electrophoresis systems, plus temperature verification for thermal cyclers and real-time PCR instruments. On site or remotely, with a written service report.

-

### [Genetic analyzer service](https://elis.al/genetic-analyzer-service/)

Maintenance, calibration and repair for the 3500 series, 3130, 3730, SeqStudio and SeqStudio Flex.

-

### [Online array regeneration](https://elis.al/regenerated-capillary-arrays/#online)

A washing liquid and a new RFID tag, with our engineer online or with our protocol.

-

### [PCR and qPCR instrument service](https://elis.al/pcr-qpcr-service/)

Block temperature verification, optical calibrations and maintenance.

### How a service request works

1.

#### Send the model and the symptom

Serial number, software version and a screenshot of the error or the electropherogram help us prepare.

2.

#### Remote diagnosis

We review logs and data with you and confirm what needs adjusting or replacing.

3.

#### Repair on site or guided

An engineer visits, or we guide your team step by step for simple jobs.

4.

#### Verification and report

Calibration and a test run confirm the result. You receive a report with the values measured.

## Which consumables fit your instrument?

The capillary count and array length decide which array you need, and the polymer format changes between instrument families. The [full compatibility chart](https://elis.al/guides/compatibility-chart/) adds buffers and every OEM part number.

**Capillary arrays and polymer by instrument**

| Instrument | Capillaries | Capillary arrays | Polymer |
|---|---|---|---|
| 3500 Genetic Analyzer | 8 | 36 cm `4404683` / 50 cm `4404685` | POP-4, POP-6, POP-7 pouches |
| 3500xL Genetic Analyzer | 24 | 36 cm `4404687` / 50 cm `4404689` | POP-4, POP-6, POP-7 pouches |
| SeqStudio 8 Flex Genetic Analyzer | 8 | 36 cm `A49104` / 50 cm `A49106` | Same as 3500 series |
| SeqStudio 24 Flex Genetic Analyzer | 24 | 36 cm `A49105` / 50 cm `A49107` | Same as 3500 series |
| SeqStudio Genetic Analyzer | 4 | Built into the SeqStudio Cartridge v2 (28 cm) | POP-1 inside the cartridge / `A41331` |
| 3730 DNA Analyzer | 48 | 36 cm `4331247` / 50 cm `4331250` | POP-7 bottle, 28 mL / `4363929` |
| 3730xl DNA Analyzer | 96 | 36 cm `4331244` / 50 cm `4331246` | POP-7 bottle, 28 mL / `4363929` |
| 3130 / 3100-Avant Genetic Analyzer | 4 | 36 cm `4333464` / 50 cm `4333466` | POP-4, POP-6, POP-7 bottles |
| 3130xl / 3100 Genetic Analyzer | 16 | 36 cm `4315931` / 50 cm `4315930` | POP-4, POP-6, POP-7 bottles |

OEM part numbers are shown for cross-reference only, so you can match what your instrument uses today.

## Practical guides for capillary electrophoresis

Written for the people who run the instrument every day, and free to print and share in your lab.

-

### [Compatibility chart](https://elis.al/guides/compatibility-chart/)

Arrays, polymers and buffers for each Applied Biosystems genetic analyzer, with OEM part numbers.

-

### [3500 troubleshooting guide](https://elis.al/guides/3500-troubleshooting/)

Likely causes and fixes for spikes, weak signal, broad peaks, pull-up, current errors and failed calibrations.

-

### [Maintenance checklist](https://elis.al/guides/maintenance-checklist/)

A printable schedule of daily, 14-day, monthly and yearly tasks for the 3500 and 3500xL.

## Supplying laboratories in four regions

We deliver consumables and arrange service for laboratories in Europe, Asia, the Middle East and Latin America. Tell us your country with the request and we confirm the delivery time and shipping cost.

[How we work](https://elis.al/about/) explains what to expect from us.

- Europe
- Asia
- Middle East
- Latin America

## Questions labs ask us

### Which capillary array do I need for a 3500 or 3500xL?

The 3500 uses an 8-capillary array and the 3500xL a 24-capillary array. Both come in 36 cm and 50 cm. Choose the length your run modules are set up for: 36 cm for fast runs and most fragment analysis, 50 cm for longer sequencing reads.

### What is the difference between POP-7, POP-6 and POP-4?

POP-7 is the general-purpose polymer for sequencing and fragment analysis. POP-6 gives higher resolution for sequencing, with longer run times. POP-4 is optimized for human identification (HID) fragment analysis.

### How long can polymer stay on a 3500?

Up to 14 days at or below 25 °C. Above 25 °C, POP-7 should be replaced after 7 days. The instrument also tracks the injection limit of each pouch and asks for a new one when either limit is reached.

### Do you list OEM part numbers?

Yes. Every product shows the matching OEM part number for cross-reference, for example 4404685 for the 3500 8-capillary 50 cm array or 4393714 for the 960-sample POP-7 pouch.

### What is a regenerated capillary array?

An original array that has been used and then cleaned, regenerated and, for the 3500 series, fitted with a reprogrammed RFID tag. It costs much less than a new array. See [regenerated capillary arrays](https://elis.al/regenerated-capillary-arrays/).

### Which instruments do you supply and service?

The 3500 and 3500xL with their Dx versions, the 3130 and 3130xl, the 3730 and 3730xl, the SeqStudio Flex and SeqStudio, and thermal cyclers and real-time PCR systems.

### Can I regenerate the arrays I already have?

Yes, for the 3130, 3130xl, 3730, 3730xl, 3500, 3500xL, 3500 Dx, 3500xL Dx and SeqStudio Flex. We send a washing liquid and, where the instrument reads RFID tags, a new tag; our engineers guide you online, or you follow our protocol.

### How do I request a quote?

Use the [quote form](https://elis.al/contact/#quote) or email us the instrument model, the items and the quantities. We reply with price, availability and delivery time.